Lascar EL-WiFi-VACX High-Accuracy WiFI Vaccine Monitoring Kit With Alarm Warning Light and Sounder

Change Currency

SKU
XE-4058-001
£227.94 £189.95
Lead Time:
5 working days
Weight:
300g
Tariff Code:
90330000
Manufacturer:
Lascar
Country of Origin:
GB
RoHS Compliant:

Additional probes available here

The EL-WiFi-VACX, with its built-in sounder and LED light-up alert functions, allows remote vaccine monitoring and data collection over a WiFi network, viewable from any internet-enabled device including smartphone, tablet or PC, via the EasyLog Cloud. Set-up high and low alarms and receive instant email alerts when temperatures become too hot or cold in the fridge or freezer.

A Glycol-filled temperature probe is supplied with the product, which matches the temperature response of vaccines in storage, ensuring highly accurate and representative readings. The probe is replaceable and can be calibrated separately from the main device.

The EL-WiFi-VACX stores data internally if it loses WiFi connection and automatically uploads it to the Cloud once reconnected.

Use the EasyLog Cloud Apps for Android and Apple to easily set up your device and access your data anytime, anywhere.

This product meets and exceeds CDC standards for vaccine monitoring.

A high accuracy Glycol-filled temperature probe, traceable calibration certificate and a wall mounting bracket are included with the product.

The EL-WiFi-VACX temperature data logger is suitable for a wide variety of applications. These include:

  • Vaccine Storage
  • Blood & Blood Component Storage
  • Pharmaceutical Cold Chain

Specifications

ModelEL-WiFi-VACX
TypeHigh-Accuracy WiFI Vaccine Monitoring Kit With Alarm Warning Light and Sounder
Communications ProtocolWiFi
Accuracy(typical) +/-0.1°C (-10 to +60°C) +/0.2°F (+14 to +140°F)
Readingsunlimited
ResolutionMeasurement resolution: 0.01°C (0.02°F) / Display resolution: 0.01
Logging IntervalUser selectablebetween 10 seconds and 12 hours
BatteryBattery Life: > 6 months
Dimensions82 x 70 x 36 mm
Temperature Range(probe supplied) -40 to +60°C (-40 to 140°F)
Operating Temperature-20 to +60°C (-4 to +140°F)
CalibrationCalibration Certificate Supplied at 2 and 8°C
Standards MetLogger IP42Probe IP67